Limiting Factors Analysis in Validation of a CryoShipper & a CryoExtra Freezer for Cryo-samples Related to the Abu Dhabi Bone Marrow Transplantation Program

Received: 25 April 2022    Accepted: 9 May 2022    Published: 31 May 2022
Views:       Downloads:
Abstract

Introduction: CryoShipper and CryoFreezer tanks were designed to safely transport and store biological material at cryogenic temperatures. An adverse warming event may occur if the CryoShipper is tilted during transit if temperature and proper levels of liquid nitrogen (LN2) vapors are not maintained. Storing cryogenic hematopoietic stem cell samples for a short or long time requires validating all pieces of equipment employed in a bone marrow transplant program. Objective: To determine the limitations of the MVE CryoShipper CX and the Thermo Fisher CryoExtra™ Cryogenic tank for the Abu Dhabi Bone Marrow Transplantation Program. Methods: The MVE CryoShipper CX was weighted and primed with LN2, and the excess was removed. After stabilizing the temperature, the shipper was tilted onto its side for 24, 72, and 96 hours. The temperature was monitored and recorded at 15 minutes intervals using a Datalogger. After returning to the correct upright position, the dry shipper temperature was also observed for five days after refilling with LN2 to evaluate the secure timing of a shipment. For an LN2 filled CryoExtra tank, dual control of the LN2 level and the temperature was done using manual measurement and automatic display. Later, vials and a bag from three healthy donors’ frozen white blood cell buffy coat samples were kept inside and defrosted daily to measure total mononuclear cell counts CD45+/7AAD cells viability by flow cytometry for five days. Results: The CryoShipper maintained cryogenic temperatures below -150°C for the entire duration of each analysis. A maximum temperature of -182.7°C was reached during the 24-hour tilt experiment from the temperature probe. The CryoExtra 0140 freezer temperature was always between -184°C / -194.2°C, and LN2 levels coincided in both measuring methods during the time slot. Stored and defrosted cells keep their % of viability and absolute number over the expected reference range and compare mean+/-standard deviation (SD) between them without assessing the statistical difference with p≥0.05. The defrosted cells’ mean viability of 83.12%, SD±9.04, and a mean of 1,781 cell/µL, SD±1,215. Conclusion: Our modern dry shipper controls much of the vapor within the shipper. The CryoShipper can withstand being tilted during transit for 96 hours without risks of an adverse warming event. The CryoExtra 0140 tank its performance was under the established parameters. Both types of cryogenic devices can be used for the safe cryopreservation of cell samples keeping all the security measures and controls as advised.

Published in Clinical Medicine Research (Volume 11, Issue 3)
DOI 10.11648/j.cmr.20221103.12
Page(s) 42-47
Creative Commons

This is an Open Access article, distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution and reproduction in any medium or format, provided the original work is properly cited.

Copyright

Copyright © The Author(s), 2024. Published by Science Publishing Group

Keywords

Cryogenic, Control Rate Freezers, Low-Temperature Storage, Safety, Thermal Risk, Validation, Liquid Nitrogen

References
[1] Niederwieser D, Baldomero H, Bazuaye N, Bupp C, Chaudhri N, Corbacioglu S, et al. (2021). Haematologica. One and a half million hematopoietic stem cell transplants: continuous and differential improvement in worldwide access with the use of non-identical family donors. DOI: 10.3324/HAEMATOL.2021.279189. ISSN: 0390-6078.
[2] Lambert JL. (2020). State of the Industry: The Financial, Clinical, and Scientific Landscape for Cell and Gene Therapies. Cell and Gene Therapy Insights. 6 (1). DOI: 10.18609/cgti.2020.007. ISSN: 20597800.
[3] U.S. Department of Health and Human Services, Food and Drug Administration, Center for Drug Evaluation and Research (CDER), Center for Biologics Evaluation and Research (CBER), Center for Veterinary Medicine (CVM). (2011). Guidance for Industry Process Validation: General Principles and Practices Guidance for Industry. 800–35. ISSN: 2402769300.
[4] FACT-JACIE. (2021). FACT-JACIE International Standards for Hematopoietic Cellular Therapy. Product Collection, Processing, and Administration EIGHTH EDITION 8.1.
[5] Villegas-Valverde CA, Ventura-Carmenate Y, M Alkaabi F, Bencomo-Hernández A. (2021). Early immune reconstitution assessment by flow/mass cytometry in autologous hematopoietic transplantation performed in a patient with multiple myeloma. Hematology & Transfusion International Journal [Internet]. [cited 2022 Mar 27]; 9 (2): 21–6. Available from: https://medcraveonline.com/HTIJ/HTIJ-09-00246.php
[6] Castillo-Aleman YM, Ventura-Carmenate Y, Rivero-Jimenez RA, Bencomo-Hernandez AA (2021) Regulatory Prospects of Clinical Trials with Stem Cells in the United Arab Emirates. Int J Stem Cell Res Ther 8: 073. doi.org/10.23937/2469-570X/1410073.
[7] Ventura-Carmenate Y, Othman J, Thabet M, Amin A, Martinez M, Kistan J, et al. (2021). P-126 | Implementation of Amicus BlueTM system in routine use for autologous mononuclear cell collection in Abu Dhabi. Vox Sanguinis [Internet]. [cited 2022 Mar 27]; 116 (suppl. 1): 86. Available from: https://www.researchgate.net/publication/352374210
[8] American Society of Testing Materials (ASTM). (2014). ASTM-D4169 - Standard Practice for Performance Testing of Shipping Container. Life Science Outsourcing, Inc. [Internet]. https://lso-inc.com/services/medical-package-testing/standards/astm-standards/astm-d4169/. [cited 2022 Mar 27]. Available from: https://lso-inc.com/services/medical-package-testing/standards/astm-standards/astm-d4169/
[9] Chart Industries Inc. (2019). MVE Vapor Shipper Series [Internet]. https://files.chartindustries.com/VaporShipperSeries_MLAI0021.pdf. [cited 2022 Mar 27]. Available from: https://files.chartindustries.com/VaporShipperSeries_MLAI0021.pdf
[10] Tran C-A, Gee AP. (2022). Transport and Shipment of Cellular and Gene Therapy Products [Internet]. Cell Therapy. Springer, Cham; [cited 2022 Mar 27]. 503–516 p. Available from: https://link.springer.com/chapter/10.1007/978-3-030-75537-9_32
[11] Thermo Fisher Scientific Inc. (2014) Thermo Scientific CryoExtra High-Efficiency Cryogenic Storage Document Connect [Internet]. [cited 2022 Mar 27]. Available from: https://www.thermofisher.com/document-connect/document-connect.html?url=https%3A%2F%2Fassets.thermofisher.com%2FTFS-Assets%2FLED%2Fbrochures%2FCryoExtra-Bro-0414.pdf
[12] Thermo Fisher Scientific Inc. (2020). Model 7450 and 7470 Series CryoMed Controlled Rate Freezer* [Internet]. [cited 2022 Mar 27]. Available from: www.thermoscientific.com/labwarranty
[13] CDC, NIH. (2020). Biosafety in Microbiological and Biomedical Laboratories. 6th Edition. Centers for Disease Control and Prevention. National Institutes of Health.
[14] World Medical Association (WMA). (2013). World Medical Association Declaration of Helsinki Ethical Principles for Medical Research Involving Human Subjects. [Internet]. 2013; Available from: www.jama.com.
[15] AERAS. Charging a Dry Shipper Training Video - YouTube [Internet]. (2014). [cited 2022 Mar 27]. Available from: https://www.youtube.com/watch?v=6wNbk3gTJHA
[16] United States Department of Labor. (2012). - Hazard Communication. | Occupational Safety and Health Administration [Internet]. 1910.1200. [cited 2022 Mar 27]. Available from: https://www.osha.gov/laws-regs/regulations/standardnumber/1910/1910.1200
[17] Elpro.com. Technical Specifications. Libero CONFIG Version 2020.7.1.0. [cited 2022 Mar 27]; Available from: www.elpro.com/downloads.
[18] Kotula PL, Areman EM, Sacher RA. (1998). Hematopoietic Stem Cell Processing and Transplantation. Red Cell Transfusion [Internet]. [cited 2022 Mar 27]; 121–37. Available from: https://link.springer.com/chapter/10.1007/978-1-4612-1798-5_8
[19] Lecoeur H, de Oliveira-Pinto LM, Gougeon ML. (2002). Multiparametric flow cytometric analysis of biochemical and functional events associated with apoptosis and oncosis using the 7-aminoactinomycin D assay. Journal of Immunological Methods. 265 (1–2): 81–96.
[20] International Stardarization Organization (ISO). (2012). ISO 15189: 2012 - Medical laboratories — Requirements for quality and competence [Internet]. [cited 2022 Mar 27]. Available from: https://www.iso.org/standard/56115.html
[21] GraphPad LLC. GraphPad Software [Internet]. web site. (2018). [cited 2022 Mar 27]. Available from: https://www.graphpad.com/quickcalcs/
[22] World Health Organization. (2020) Handbook Good Laboratory Practice (GLP). Quality practices for regulated non-clinical research and development. Special Programme for Research & Training in Tropical Diseases (TDR) sponsored by UNICEF/UNDP/WorldBank/WHOSECONDEDITION.
[23] Berry L, Horton R, Leonforte C, Gibson D. (2019). MVE CryoShipper CT-50 Tilt Validation for Cryogenic Cord Blood Shipments. Stem Cells Translational Medicine [Internet]. [cited 2022 Mar 27]; 8 (S1): S35–S35. Available from: https://academic.oup.com/stcltm/article/8/S1/S35/6403864
[24] Simione F, Sharp T. (2017) Best practices for storing and shipping cryopreserved cells. In Vitro Cellular and Developmental Biology - Animal. 53 (10): 888–95.
Cite This Article
  • APA Style

    Rene Antonio Rivero-Jimenez, Zain Asif, Muhammad Touseef Haider, Yandy Marx Castillo-Aleman, Yendry Ventura-Carmenate, et al. (2022). Limiting Factors Analysis in Validation of a CryoShipper & a CryoExtra Freezer for Cryo-samples Related to the Abu Dhabi Bone Marrow Transplantation Program. Clinical Medicine Research, 11(3), 42-47. https://doi.org/10.11648/j.cmr.20221103.12

    Copy | Download

    ACS Style

    Rene Antonio Rivero-Jimenez; Zain Asif; Muhammad Touseef Haider; Yandy Marx Castillo-Aleman; Yendry Ventura-Carmenate, et al. Limiting Factors Analysis in Validation of a CryoShipper & a CryoExtra Freezer for Cryo-samples Related to the Abu Dhabi Bone Marrow Transplantation Program. Clin. Med. Res. 2022, 11(3), 42-47. doi: 10.11648/j.cmr.20221103.12

    Copy | Download

    AMA Style

    Rene Antonio Rivero-Jimenez, Zain Asif, Muhammad Touseef Haider, Yandy Marx Castillo-Aleman, Yendry Ventura-Carmenate, et al. Limiting Factors Analysis in Validation of a CryoShipper & a CryoExtra Freezer for Cryo-samples Related to the Abu Dhabi Bone Marrow Transplantation Program. Clin Med Res. 2022;11(3):42-47. doi: 10.11648/j.cmr.20221103.12

    Copy | Download

  • @article{10.11648/j.cmr.20221103.12,
      author = {Rene Antonio Rivero-Jimenez and Zain Asif and Muhammad Touseef Haider and Yandy Marx Castillo-Aleman and Yendry Ventura-Carmenate and Carlos Agustin Villegas-Valverde and Antonio Alfonso Bencomo-Hernandez},
      title = {Limiting Factors Analysis in Validation of a CryoShipper & a CryoExtra Freezer for Cryo-samples Related to the Abu Dhabi Bone Marrow Transplantation Program},
      journal = {Clinical Medicine Research},
      volume = {11},
      number = {3},
      pages = {42-47},
      doi = {10.11648/j.cmr.20221103.12},
      url = {https://doi.org/10.11648/j.cmr.20221103.12},
      eprint = {https://article.sciencepublishinggroup.com/pdf/10.11648.j.cmr.20221103.12},
      abstract = {Introduction: CryoShipper and CryoFreezer tanks were designed to safely transport and store biological material at cryogenic temperatures. An adverse warming event may occur if the CryoShipper is tilted during transit if temperature and proper levels of liquid nitrogen (LN2) vapors are not maintained. Storing cryogenic hematopoietic stem cell samples for a short or long time requires validating all pieces of equipment employed in a bone marrow transplant program. Objective: To determine the limitations of the MVE CryoShipper CX and the Thermo Fisher CryoExtra™ Cryogenic tank for the Abu Dhabi Bone Marrow Transplantation Program. Methods: The MVE CryoShipper CX was weighted and primed with LN2, and the excess was removed. After stabilizing the temperature, the shipper was tilted onto its side for 24, 72, and 96 hours. The temperature was monitored and recorded at 15 minutes intervals using a Datalogger. After returning to the correct upright position, the dry shipper temperature was also observed for five days after refilling with LN2 to evaluate the secure timing of a shipment. For an LN2 filled CryoExtra tank, dual control of the LN2 level and the temperature was done using manual measurement and automatic display. Later, vials and a bag from three healthy donors’ frozen white blood cell buffy coat samples were kept inside and defrosted daily to measure total mononuclear cell counts CD45+/7AAD cells viability by flow cytometry for five days. Results: The CryoShipper maintained cryogenic temperatures below -150°C for the entire duration of each analysis. A maximum temperature of -182.7°C was reached during the 24-hour tilt experiment from the temperature probe. The CryoExtra 0140 freezer temperature was always between -184°C / -194.2°C, and LN2 levels coincided in both measuring methods during the time slot. Stored and defrosted cells keep their % of viability and absolute number over the expected reference range and compare mean+/-standard deviation (SD) between them without assessing the statistical difference with p≥0.05. The defrosted cells’ mean viability of 83.12%, SD±9.04, and a mean of 1,781 cell/µL, SD±1,215. Conclusion: Our modern dry shipper controls much of the vapor within the shipper. The CryoShipper can withstand being tilted during transit for 96 hours without risks of an adverse warming event. The CryoExtra 0140 tank its performance was under the established parameters. Both types of cryogenic devices can be used for the safe cryopreservation of cell samples keeping all the security measures and controls as advised.},
     year = {2022}
    }
    

    Copy | Download

  • TY  - JOUR
    T1  - Limiting Factors Analysis in Validation of a CryoShipper & a CryoExtra Freezer for Cryo-samples Related to the Abu Dhabi Bone Marrow Transplantation Program
    AU  - Rene Antonio Rivero-Jimenez
    AU  - Zain Asif
    AU  - Muhammad Touseef Haider
    AU  - Yandy Marx Castillo-Aleman
    AU  - Yendry Ventura-Carmenate
    AU  - Carlos Agustin Villegas-Valverde
    AU  - Antonio Alfonso Bencomo-Hernandez
    Y1  - 2022/05/31
    PY  - 2022
    N1  - https://doi.org/10.11648/j.cmr.20221103.12
    DO  - 10.11648/j.cmr.20221103.12
    T2  - Clinical Medicine Research
    JF  - Clinical Medicine Research
    JO  - Clinical Medicine Research
    SP  - 42
    EP  - 47
    PB  - Science Publishing Group
    SN  - 2326-9057
    UR  - https://doi.org/10.11648/j.cmr.20221103.12
    AB  - Introduction: CryoShipper and CryoFreezer tanks were designed to safely transport and store biological material at cryogenic temperatures. An adverse warming event may occur if the CryoShipper is tilted during transit if temperature and proper levels of liquid nitrogen (LN2) vapors are not maintained. Storing cryogenic hematopoietic stem cell samples for a short or long time requires validating all pieces of equipment employed in a bone marrow transplant program. Objective: To determine the limitations of the MVE CryoShipper CX and the Thermo Fisher CryoExtra™ Cryogenic tank for the Abu Dhabi Bone Marrow Transplantation Program. Methods: The MVE CryoShipper CX was weighted and primed with LN2, and the excess was removed. After stabilizing the temperature, the shipper was tilted onto its side for 24, 72, and 96 hours. The temperature was monitored and recorded at 15 minutes intervals using a Datalogger. After returning to the correct upright position, the dry shipper temperature was also observed for five days after refilling with LN2 to evaluate the secure timing of a shipment. For an LN2 filled CryoExtra tank, dual control of the LN2 level and the temperature was done using manual measurement and automatic display. Later, vials and a bag from three healthy donors’ frozen white blood cell buffy coat samples were kept inside and defrosted daily to measure total mononuclear cell counts CD45+/7AAD cells viability by flow cytometry for five days. Results: The CryoShipper maintained cryogenic temperatures below -150°C for the entire duration of each analysis. A maximum temperature of -182.7°C was reached during the 24-hour tilt experiment from the temperature probe. The CryoExtra 0140 freezer temperature was always between -184°C / -194.2°C, and LN2 levels coincided in both measuring methods during the time slot. Stored and defrosted cells keep their % of viability and absolute number over the expected reference range and compare mean+/-standard deviation (SD) between them without assessing the statistical difference with p≥0.05. The defrosted cells’ mean viability of 83.12%, SD±9.04, and a mean of 1,781 cell/µL, SD±1,215. Conclusion: Our modern dry shipper controls much of the vapor within the shipper. The CryoShipper can withstand being tilted during transit for 96 hours without risks of an adverse warming event. The CryoExtra 0140 tank its performance was under the established parameters. Both types of cryogenic devices can be used for the safe cryopreservation of cell samples keeping all the security measures and controls as advised.
    VL  - 11
    IS  - 3
    ER  - 

    Copy | Download

Author Information
  • Advanced Technology Laboratory Department, Abu Dhabi Stem Cells Center, Abu Dhabi, the United Arab Emirates

  • Advanced Technology Laboratory Department, Abu Dhabi Stem Cells Center, Abu Dhabi, the United Arab Emirates

  • Advanced Technology Laboratory Department, Abu Dhabi Stem Cells Center, Abu Dhabi, the United Arab Emirates

  • Immunology Department, Abu Dhabi Stem Cells Center, Abu Dhabi, the United Arab Emirates

  • Immunology Department, Abu Dhabi Stem Cells Center, Abu Dhabi, the United Arab Emirates

  • Advanced Technology Laboratory Department, Abu Dhabi Stem Cells Center, Abu Dhabi, the United Arab Emirates

  • Advanced Technology Laboratory Department, Abu Dhabi Stem Cells Center, Abu Dhabi, the United Arab Emirates

  • Sections